1.1   This document specifies requirements and tests for LTSF sterilizers, which use a mixture of low temperature steam and formaldehyde as sterilizing agent, and which are working below ambient pressure only.
These sterilizers are primarily used for the sterilization of heat labile medical devices in health care facilities.
1.2   This document specifies minimum requirements:
-   for the performance and design of sterilizers intended to deliver an LTSF process capable of sterilizing medical devices;
-   for the equipment and controls of these sterilizers which are needed for operation, control and monitoring of the sterilization processes, and which can be used for validation of the sterilization process.
1.3   This document specifies further test equipment and test procedures used to verify conformance of the equipment design and performance specified by this document.
1.4   This document does not specify requirements and tests for decontamination systems for use in rooms, enclosures, or environmental spaces.

  • Standard
    127 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies general performance requirements for washer-disinfectors (WD) and washer-disinfector accessories that are intended to be used for cleaning and disinfection of reusable medical devices. It specifies performance requirements for cleaning and disinfection as well as for the accessories that can be required to achieve the necessary performance. The methods and instrumentation required for validation, routine control and monitoring and requalification, periodically and after essential repairs, are also specified.
NOTE 1        The requirements can be applied to WD intended for use with other articles used in the context of medical, dental, pharmaceutical and veterinary practice.
The requirements for WD intended to process specific loads are specified in ISO 15883-2, ISO 15883-3, ISO 15883-4, ISO 15883-6 and ISO 15883-7. For WD intended to process loads of two or more different types, the requirements of the applicable parts of ISO 15883-2, ISO 15883-3, ISO 15883-4, ISO 15883-6 and ISO 15883-7 apply.
This document does not specify requirements intended for machines for use for laundry or general catering purposes.
This document does not include requirements for machines which are intended to sterilize the load, or which are designated as “sterilizers” and addressed in other standards.
The specified performance requirements of this document do not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
NOTE 2        Chemicals in some cleaning agents and disinfectants can react with prion protein in a manner that can inhibit its removal or inactivation.  If the presence of prion protein is considered a possibility, then this can influence the choice of cleaning agent and disinfectant.
NOTE 3        This document can be used by prospective purchasers and manufacturers as the basis of agreement on the specification of a WD. The test methods for demonstration of conformity with the requirements of this document can also be employed by users to demonstrate continued conformity of the installed WD throughout its service life. Guidance on a routine test programme is given in Annex A.

  • Standard
    91 pages
    English language
    sale 10% off
    e-Library read for
    1 day

1.1   This document specifies requirements and tests for LTSF sterilizers, which use a mixture of low temperature steam and formaldehyde as sterilizing agent, and which are working below ambient pressure only.
These sterilizers are primarily used for the sterilization of heat labile medical devices in health care facilities.
1.2   This document specifies minimum requirements:
-   for the performance and design of sterilizers intended to deliver an LTSF process capable of sterilizing medical devices;
-   for the equipment and controls of these sterilizers which are needed for operation, control and monitoring of the sterilization processes, and which can be used for validation of the sterilization process.
1.3   This document specifies further test equipment and test procedures used to verify conformance of the equipment design and performance specified by this document.
1.4   This document does not specify requirements and tests for decontamination systems for use in rooms, enclosures, or environmental spaces.

  • Standard
    127 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies general performance requirements for washer-disinfectors (WD) and washer-disinfector accessories that are intended to be used for cleaning and disinfection of reusable medical devices. It specifies performance requirements for cleaning and disinfection as well as for the accessories that can be required to achieve the necessary performance. The methods and instrumentation required for validation, routine control and monitoring and requalification, periodically and after essential repairs, are also specified.
NOTE 1        The requirements can be applied to WD intended for use with other articles used in the context of medical, dental, pharmaceutical and veterinary practice.
The requirements for WD intended to process specific loads are specified in ISO 15883-2, ISO 15883-3, ISO 15883-4, ISO 15883-6 and ISO 15883-7. For WD intended to process loads of two or more different types, the requirements of the applicable parts of ISO 15883-2, ISO 15883-3, ISO 15883-4, ISO 15883-6 and ISO 15883-7 apply.
This document does not specify requirements intended for machines for use for laundry or general catering purposes.
This document does not include requirements for machines which are intended to sterilize the load, or which are designated as “sterilizers” and addressed in other standards.
The specified performance requirements of this document do not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
NOTE 2        Chemicals in some cleaning agents and disinfectants can react with prion protein in a manner that can inhibit its removal or inactivation.  If the presence of prion protein is considered a possibility, then this can influence the choice of cleaning agent and disinfectant.
NOTE 3        This document can be used by prospective purchasers and manufacturers as the basis of agreement on the specification of a WD. The test methods for demonstration of conformity with the requirements of this document can also be employed by users to demonstrate continued conformity of the installed WD throughout its service life. Guidance on a routine test programme is given in Annex A.

  • Standard
    91 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document addresses good clinical practice for the design, conduct, recording and reporting of clinical investigations carried out in
human subjects to assess the clinical performance or effectiveness and safety of medical devices.
For post-market clinical investigations, the principles set forth in this document are intended to be followed as far as relevant,
considering the nature of the clinical investigation (see AnnexI).
This document specifies general requirements intended to
- protect the rights, safety and well-being of human subjects,
- ensure the scientific conduct of the clinical investigation and the credibility of the clinical investigation results,
- define the responsibilities of the sponsor and principal investigator, and
- assist sponsors, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of
medical devices.
NOTE1 Users of this document need to consider whether other standards and/or national requirements also apply to the
investigational device(s) under consideration or the clinical investigation. If differences in requirements exist, the most stringent apply.
NOTE2 For Software as a Medical Device (SaMD) demonstration of the analytical validity (the SaMD’s output is accurate for a given
input), and where appropriate, the scientific validity (the SaMD’s output is associated to the intended clinical condition/physiological
state), and clinical performance (the SaMD’s output yields a clinically meaningful association to the target use) of the SaMD, the
requirements of this document apply as far as relevant (see Reference[4]). Justifications for exemptions from this document can
consider the uniqueness of indirect contact between subjects and the SaMD.
This document does not apply to in vitro diagnostic medical devices. However, there can be situations, dependent on the device and
national or regional requirements, whereusers of this document might consider whether specific sections and/or requirements of this
document could be applicable.

  • Amendment
    13 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies general requirements for non-active surgical implants, hereafter referred to as implants.
This document is not applicable to dental implants, dental restorative materials, transendodontic and transradicular implants, intra-ocular lenses and implants utilizing viable animal or human tissue.
With regard to safety, this document specifies requirements for intended performance, design attributes, materials, design evaluation, manufacture, sterilization, packaging and information supplied by the manufacturer, and tests to demonstrate compliance with these requirements.
Additional requirements applicable to specific implants or implant families are given or referred to in Level 2 and Level 3 standards.
NOTE 1        This document does not require that the manufacturer have a quality management system in place. However, many regulatory authorities require the application of a quality management system, such as that described in ISO 13485, to ensure that the implant achieves its intended performance and safety.
NOTE 2        In this document, when not otherwise specified, the term "implant" refers to each individual component of a system or a modular implant, provided separately or as a set of components, as well as to all ancillary implants or associated implants designed for improving the intended performance.

  • Standard
    36 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies general requirements for non-active surgical implants, hereafter referred to as implants.
This document is not applicable to dental implants, dental restorative materials, transendodontic and transradicular implants, intra-ocular lenses and implants utilizing viable animal or human tissue.
With regard to safety, this document specifies requirements for intended performance, design attributes, materials, design evaluation, manufacture, sterilization, packaging and information supplied by the manufacturer, and tests to demonstrate compliance with these requirements.
Additional requirements applicable to specific implants or implant families are given or referred to in Level 2 and Level 3 standards.
NOTE 1        This document does not require that the manufacturer have a quality management system in place. However, many regulatory authorities require the application of a quality management system, such as that described in ISO 13485, to ensure that the implant achieves its intended performance and safety.
NOTE 2        In this document, when not otherwise specified, the term "implant" refers to each individual component of a system or a modular implant, provided separately or as a set of components, as well as to all ancillary implants or associated implants designed for improving the intended performance.

  • Standard
    36 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document addresses good clinical practice for the design, conduct, recording and reporting of clinical investigations carried out in
human subjects to assess the clinical performance or effectiveness and safety of medical devices.
For post-market clinical investigations, the principles set forth in this document are intended to be followed as far as relevant,
considering the nature of the clinical investigation (see AnnexI).
This document specifies general requirements intended to
- protect the rights, safety and well-being of human subjects,
- ensure the scientific conduct of the clinical investigation and the credibility of the clinical investigation results,
- define the responsibilities of the sponsor and principal investigator, and
- assist sponsors, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of
medical devices.
NOTE1 Users of this document need to consider whether other standards and/or national requirements also apply to the
investigational device(s) under consideration or the clinical investigation. If differences in requirements exist, the most stringent apply.
NOTE2 For Software as a Medical Device (SaMD) demonstration of the analytical validity (the SaMD’s output is accurate for a given
input), and where appropriate, the scientific validity (the SaMD’s output is associated to the intended clinical condition/physiological
state), and clinical performance (the SaMD’s output yields a clinically meaningful association to the target use) of the SaMD, the
requirements of this document apply as far as relevant (see Reference[4]). Justifications for exemptions from this document can
consider the uniqueness of indirect contact between subjects and the SaMD.
This document does not apply to in vitro diagnostic medical devices. However, there can be situations, dependent on the device and
national or regional requirements, whereusers of this document might consider whether specific sections and/or requirements of this
document could be applicable.

  • Amendment
    13 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the performance requirements for sterile, single-use non-active hydrocephalus shunts. This includes not only the valve, but also additional components such as tubes and reservoirs.
This document does not provide any recommendations on which type of valve is most suitable for any specific context of use.
This document specifies the mechanical and technical requirements to manufacture shunts and the technical information of the valve to be supplied by the manufacturer.
This document does not apply to active implants for the treatment of hydrocephalus.

  • Standard
    18 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the performance requirements for sterile, single-use non-active hydrocephalus shunts. This includes not only the valve, but also additional components such as tubes and reservoirs.
This document does not provide any recommendations on which type of valve is most suitable for any specific context of use.
This document specifies the mechanical and technical requirements to manufacture shunts and the technical information of the valve to be supplied by the manufacturer.
This document does not apply to active implants for the treatment of hydrocephalus.

  • Standard
    18 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies:
—     the general principles governing the biological evaluation within a risk management process of the gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments;
—     the general categorization of gas pathways based on the nature and duration of their contact with the gas stream;
—     the evaluation of existing relevant data from all sources;
—     the identification of gaps in the available data set on the basis of a risk analysis;
—     the identification of additional data sets necessary to analyse the biological safety of the gas pathway;
—     the assessment of the biological safety of the gas pathway.
This document covers general principles regarding biocompatibility assessment of medical device materials, which make up the gas pathway, in normal use and normal condition. This document does not cover biological hazards arising from mechanical damage.
The other parts of ISO 18562 cover specific tests that address potentially hazardous substances that are added to the respirable gas stream and establish acceptance criteria for these substances.
This document addresses potential contamination of the gas stream arising from the gas pathways within the medical device, which might then be conducted to the patient.
This document applies over the expected lifetime of the medical device when operated according to the instructions for use. This includes degradation arising from exposure to environmental conditions as well as cleaning, disinfection and sterilisation (i.e. processing). It also includes user action or inaction (omission) that leads to an unintended or unexpected outcome (result) (i.e. use error). It does not include conscious/intentional action or inaction that violates the instructions for use and is beyond reasonable risk control by the manufacturer (i.e. abnormal use).
This document does not address biological evaluation of the surfaces of medical devices that have direct contact with the patient or user. The requirements for direct contact surfaces are found in the ISO 10993 series.
Medical devices, parts or accessories containing gas pathways that are addressed by this document include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving equipment, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing system filters and Y-pieces as well as any breathing accessories intended to be used with such medical devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document.
This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use.
EXAMPLE            Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder, or room air taken into the medical device is not addressed by ISO 18562 (all parts).

  • Standard
    49 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies tests for the emissions of particulate matter from the gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments. The tests of this document are intended to quantify particles from 0,25 µm diameter to 10 µm diameter that are emitted by the medical device, its parts or accessories into the respirable gas stream. This document establishes acceptance criteria for these tests.
This document does not address nanoparticles. Insufficient data exist to establish exposure limits for particles less than 0,25 µm diameter.
This document does not address particles larger than 10 µm diameter. These particles are deposited in the nasal cavity. Additional information can be needed for medical devices or accessories that bypass the nose. This is outside the scope of this document but can be required by some authorities having jurisdiction.
This document therefore adopts the same approach as the US Environmental Protection Agency (EPA) in setting limits based solely on particle size and not their chemistry.
This document addresses potential contamination of the gas stream arising from the gas pathways, which is then conducted to the patient.
This document applies over the expected lifetime of the medical device in normal use and takes into account the effects of any intended processing.
This document does not address biological evaluation of the particles that are deliberately released by a nebulizer (i.e. the therapeutic agent).
This document does not address biological evaluation of the surfaces of gas pathways that have direct contact with the patient. The requirements for direct contact surfaces are found in the ISO 10993 series.
Medical devices, parts or accessories, containing gas pathways that are addressed by this document, include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing systems filters, Y-pieces, and any breathing accessories intended to be used with such devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document.
This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use.
EXAMPLE            Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder or room air taken into the medical device is not addressed by ISO 18562 (all parts).

  • Standard
    27 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies tests for the emissions of volatile organic substances from the gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments. The tests of this document are intended to quantify emissions of volatile organic substances that are added to the respirable gas stream by the materials of the gas pathway. This document establishes acceptance criteria for these tests.
NOTE            Gaseous emission of volatile organic substances includes emissions of volatile organic compounds, semi-volatile organic compounds and very volatile organic compounds.
This document addresses potential contamination of the gas stream arising from the gas pathways of medical devices or accessories, which is then conducted to the patient.
This document applies over the expected lifetime of the medical device in normal use and takes into account the effects of any intended processing.
This document does not address biological evaluation of the surfaces of gas pathways that are in direct contact with the patient. The requirements for direct contact surfaces are found in the ISO 10993 series.
Medical devices, parts or accessories containing gas pathways that are addressed by this document include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing systems filters, Y-pieces and any breathing accessories intended to be used with such devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document.
This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use.
EXAMPLE            Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder or room air taken into the medical device is not addressed by ISO 18562 series.
This document is intended to be read in conjunction with ISO 18562-1.

  • Standard
    27 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies tests for substances leached by liquid water condensing in gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments. The chemical characterization methods described in this document apply to chemical substances that could leach from the medical device, its parts or accessories into the condensate. This document establishes verifiable acceptance criteria for these tests. The identity and quantity of each chemical released is intended for toxicological risk assessment as described in ISO 18562-1:2024.
This document addresses potential contamination of the gas stream arising from the gas pathways, which deliver breathing gas to the patient.
This document applies over the expected lifetime of the medical device in normal use and takes into account the effects of any intended processing.
This document does not address biological evaluation of the surfaces of gas pathways that have direct contact with the patient. The requirements for direct contact surfaces are found in the ISO 10993 series.
Medical devices, parts or accessories containing gas pathways that are addressed by this document include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing systems filters, Y-pieces and any breathing accessories intended to be used with such devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document.
This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use.
EXAMPLE            Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder, or room air taken into the medical device.
This document does not address contact with drugs or anaesthetic agents. If a medical device or accessory is intended to be used with anaesthetic agents or drugs, then additional testing can be required. This document is intended to quantify hazardous water-soluble substances that are leached from the medical device, its parts or accessories by condensate and then conveyed by that liquid to the patient.

  • Standard
    29 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies tests for substances leached by liquid water condensing in gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments. The chemical characterization methods described in this document apply to chemical substances that could leach from the medical device, its parts or accessories into the condensate. This document establishes verifiable acceptance criteria for these tests. The identity and quantity of each chemical released is intended for toxicological risk assessment as described in ISO 18562-1:2024.
This document addresses potential contamination of the gas stream arising from the gas pathways, which deliver breathing gas to the patient.
This document applies over the expected lifetime of the medical device in normal use and takes into account the effects of any intended processing.
This document does not address biological evaluation of the surfaces of gas pathways that have direct contact with the patient. The requirements for direct contact surfaces are found in the ISO 10993 series.
Medical devices, parts or accessories containing gas pathways that are addressed by this document include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing systems filters, Y-pieces and any breathing accessories intended to be used with such devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document.
This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use.
EXAMPLE            Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder, or room air taken into the medical device.
This document does not address contact with drugs or anaesthetic agents. If a medical device or accessory is intended to be used with anaesthetic agents or drugs, then additional testing can be required. This document is intended to quantify hazardous water-soluble substances that are leached from the medical device, its parts or accessories by condensate and then conveyed by that liquid to the patient.

  • Standard
    29 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies:
—     the general principles governing the biological evaluation within a risk management process of the gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments;
—     the general categorization of gas pathways based on the nature and duration of their contact with the gas stream;
—     the evaluation of existing relevant data from all sources;
—     the identification of gaps in the available data set on the basis of a risk analysis;
—     the identification of additional data sets necessary to analyse the biological safety of the gas pathway;
—     the assessment of the biological safety of the gas pathway.
This document covers general principles regarding biocompatibility assessment of medical device materials, which make up the gas pathway, in normal use and normal condition. This document does not cover biological hazards arising from mechanical damage.
The other parts of ISO 18562 cover specific tests that address potentially hazardous substances that are added to the respirable gas stream and establish acceptance criteria for these substances.
This document addresses potential contamination of the gas stream arising from the gas pathways within the medical device, which might then be conducted to the patient.
This document applies over the expected lifetime of the medical device when operated according to the instructions for use. This includes degradation arising from exposure to environmental conditions as well as cleaning, disinfection and sterilisation (i.e. processing). It also includes user action or inaction (omission) that leads to an unintended or unexpected outcome (result) (i.e. use error). It does not include conscious/intentional action or inaction that violates the instructions for use and is beyond reasonable risk control by the manufacturer (i.e. abnormal use).
This document does not address biological evaluation of the surfaces of medical devices that have direct contact with the patient or user. The requirements for direct contact surfaces are found in the ISO 10993 series.
Medical devices, parts or accessories containing gas pathways that are addressed by this document include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving equipment, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing system filters and Y-pieces as well as any breathing accessories intended to be used with such medical devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document.
This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use.
EXAMPLE            Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder, or room air taken into the medical device is not addressed by ISO 18562 (all parts).

  • Standard
    49 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies tests for the emissions of particulate matter from the gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments. The tests of this document are intended to quantify particles from 0,25 µm diameter to 10 µm diameter that are emitted by the medical device, its parts or accessories into the respirable gas stream. This document establishes acceptance criteria for these tests.
This document does not address nanoparticles. Insufficient data exist to establish exposure limits for particles less than 0,25 µm diameter.
This document does not address particles larger than 10 µm diameter. These particles are deposited in the nasal cavity. Additional information can be needed for medical devices or accessories that bypass the nose. This is outside the scope of this document but can be required by some authorities having jurisdiction.
This document therefore adopts the same approach as the US Environmental Protection Agency (EPA) in setting limits based solely on particle size and not their chemistry.
This document addresses potential contamination of the gas stream arising from the gas pathways, which is then conducted to the patient.
This document applies over the expected lifetime of the medical device in normal use and takes into account the effects of any intended processing.
This document does not address biological evaluation of the particles that are deliberately released by a nebulizer (i.e. the therapeutic agent).
This document does not address biological evaluation of the surfaces of gas pathways that have direct contact with the patient. The requirements for direct contact surfaces are found in the ISO 10993 series.
Medical devices, parts or accessories, containing gas pathways that are addressed by this document, include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing systems filters, Y-pieces, and any breathing accessories intended to be used with such devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document.
This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use.
EXAMPLE            Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder or room air taken into the medical device is not addressed by ISO 18562 (all parts).

  • Standard
    27 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies tests for the emissions of volatile organic substances from the gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments. The tests of this document are intended to quantify emissions of volatile organic substances that are added to the respirable gas stream by the materials of the gas pathway. This document establishes acceptance criteria for these tests.
NOTE            Gaseous emission of volatile organic substances includes emissions of volatile organic compounds, semi-volatile organic compounds and very volatile organic compounds.
This document addresses potential contamination of the gas stream arising from the gas pathways of medical devices or accessories, which is then conducted to the patient.
This document applies over the expected lifetime of the medical device in normal use and takes into account the effects of any intended processing.
This document does not address biological evaluation of the surfaces of gas pathways that are in direct contact with the patient. The requirements for direct contact surfaces are found in the ISO 10993 series.
Medical devices, parts or accessories containing gas pathways that are addressed by this document include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing systems filters, Y-pieces and any breathing accessories intended to be used with such devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document.
This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use.
EXAMPLE            Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder or room air taken into the medical device is not addressed by ISO 18562 series.
This document is intended to be read in conjunction with ISO 18562-1.

  • Standard
    27 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the design and dimensions for two small-bore connectors intended to be used for connections in respiratory applications of medical devices and accessories. One connector (R1) is intended for use on medical devices and accessories subjected to pressures up to 15 kPa (e.g. a breathing system). The other connector (R2) is intended for use on medical devices and accessories subjected to higher pressures between 15 kPa and 600 kPa (e.g. oxygen therapy tubing).
NOTE 1        The pressure is related to pressure available at the source to which the medical device is connected.
NOTE 2        The intended application does not preclude the use of other connectors on medical devices or accessories within this application.
NOTE 3        Requirements for alternative connectors for this intended application are specified in ISO 80369-1.
This document does not specify requirements for the medical devices or accessories that use these connectors. Such requirements are given in device-specific standards.
NOTE 4        If a device-specific standard does not exist, the performance and material requirements specified in ISO 80369-1 can be used as guidance.

  • Standard
    63 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the design and dimensions for two small-bore connectors intended to be used for connections in respiratory applications of medical devices and accessories. One connector (R1) is intended for use on medical devices and accessories subjected to pressures up to 15 kPa (e.g. a breathing system). The other connector (R2) is intended for use on medical devices and accessories subjected to higher pressures between 15 kPa and 600 kPa (e.g. oxygen therapy tubing).
NOTE 1        The pressure is related to pressure available at the source to which the medical device is connected.
NOTE 2        The intended application does not preclude the use of other connectors on medical devices or accessories within this application.
NOTE 3        Requirements for alternative connectors for this intended application are specified in ISO 80369-1.
This document does not specify requirements for the medical devices or accessories that use these connectors. Such requirements are given in device-specific standards.
NOTE 4        If a device-specific standard does not exist, the performance and material requirements specified in ISO 80369-1 can be used as guidance.

  • Standard
    63 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for hip-joint replacement implants. With regard to safety, this document specifies requirements for intended performance, design attributes, materials, design evaluation, manufacture, sterilization, packaging, information supplied by the manufacturer and methods of test.
This document applies to both total and partial hip joint replacement implants. It applies to components made of metallic and non-metallic materials.
This document applies to a wide variety of hip replacement implants, but for some specific hip replacement implant types, some considerations, not specifically covered in this document, can be applicable. Further details are given in 7.2.1.2.
The requirements which are specified in this document are not intended to require the re-design or re-testing of implants which have been legally marketed and for which there is a history of sufficient and safe clinical use. For such implants, compliance with this document can be demonstrated by providing evidence of the implant’s sufficient and safe clinical use.

  • Standard
    41 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for knee-joint replacement implants. Regarding safety, this document specifies requirements for intended performance, design attributes, materials, design evaluation, manufacture, sterilization, packaging, information supplied by the manufacturer and methods of test.
This document applies to both total and partial knee joint replacement implants. It applies to these replacements both with and without the replacement of the patella-femoral joint. It applies to components made of metallic and non-metallic materials.
This document applies to a wide variety of knee replacement implants, but for some specific knee replacement implant types, some considerations, not specifically covered in this document, can be applicable. Further details are given in 7.2.1.2.
The requirements which are specified in this document are not intended to require the re-design or re-testing of implants which have been legally marketed and for which there is a history of sufficient and safe clinical use. For such implants, compliance with this document can be demonstrated by providing evidence of the implant’s sufficient and safe clinical use.

  • Standard
    36 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for hip-joint replacement implants. With regard to safety, this document specifies requirements for intended performance, design attributes, materials, design evaluation, manufacture, sterilization, packaging, information supplied by the manufacturer and methods of test.
This document applies to both total and partial hip joint replacement implants. It applies to components made of metallic and non-metallic materials.
This document applies to a wide variety of hip replacement implants, but for some specific hip replacement implant types, some considerations, not specifically covered in this document, can be applicable. Further details are given in 7.2.1.2.
The requirements which are specified in this document are not intended to require the re-design or re-testing of implants which have been legally marketed and for which there is a history of sufficient and safe clinical use. For such implants, compliance with this document can be demonstrated by providing evidence of the implant’s sufficient and safe clinical use.

  • Standard
    41 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for knee-joint replacement implants. Regarding safety, this document specifies requirements for intended performance, design attributes, materials, design evaluation, manufacture, sterilization, packaging, information supplied by the manufacturer and methods of test.
This document applies to both total and partial knee joint replacement implants. It applies to these replacements both with and without the replacement of the patella-femoral joint. It applies to components made of metallic and non-metallic materials.
This document applies to a wide variety of knee replacement implants, but for some specific knee replacement implant types, some considerations, not specifically covered in this document, can be applicable. Further details are given in 7.2.1.2.
The requirements which are specified in this document are not intended to require the re-design or re-testing of implants which have been legally marketed and for which there is a history of sufficient and safe clinical use. For such implants, compliance with this document can be demonstrated by providing evidence of the implant’s sufficient and safe clinical use.

  • Standard
    36 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) reagents, calibrators, and controls intended for self-testing.
This document can also be applicable to accessories.
This document is applicable to the labels for outer and immediate containers and to the instructions for use.
This document does not apply to:
a) IVD instruments or equipment;
b) IVD reagents for professional use.

  • Standard
    24 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) reagents, calibrators and controls intended for professional use.
This document can also be applicable to accessories.
This document is applicable to the labels for outer and immediate containers and to the instructions for use.
This document does not apply to:
a) IVD instruments or equipment;
b) IVD reagents for self-testing.

  • Standard
    24 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) instruments intended for professional use.
This document also applies to apparatus and equipment intended to be used with IVD instruments for professional use.
This document can also be applicable to accessories.
This document does not apply to:
a) instructions for instrument servicing or repair;
b) IVD reagents, including calibrators and control materials for use in control of the reagent;
c) IVD instruments for self-testing.

  • Standard
    22 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) instruments intended for self-testing.
This document is also applicable to apparatus and equipment intended to be used with IVD instruments for self-testing.
This document can also be applicable to accessories.
This document does not apply to:
a) instructions for instrument servicing or repair;
b) IVD reagents, including calibrators and control materials for use in control of the reagent;
c) IVD instruments for professional use.

  • Standard
    21 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document defines concepts, establishes general principles, and specifies essential requirements for information supplied by the manufacturer of IVD medical devices.
This document does not address language requirements since that is the domain of national laws and regulations.
This document does not apply to:
a) IVD medical devices for performance evaluation (e.g. for investigational use only);
b) shipping documents;
c) material safety data sheets / Safety Data Sheets;
d) marketing information (consistent with applicable legal requirements).
.

  • Standard
    65 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) instruments intended for professional use.
This document also applies to apparatus and equipment intended to be used with IVD instruments for professional use.
This document can also be applicable to accessories.
This document does not apply to:
a) instructions for instrument servicing or repair;
b) IVD reagents, including calibrators and control materials for use in control of the reagent;
c) IVD instruments for self-testing.

  • Standard
    22 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document defines concepts, establishes general principles, and specifies essential requirements for information supplied by the manufacturer of IVD medical devices.
This document does not address language requirements since that is the domain of national laws and regulations.
This document does not apply to:
a) IVD medical devices for performance evaluation (e.g. for investigational use only);
b) shipping documents;
c) material safety data sheets / Safety Data Sheets;
d) marketing information (consistent with applicable legal requirements).
.

  • Standard
    65 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) reagents, calibrators and controls intended for professional use.
This document can also be applicable to accessories.
This document is applicable to the labels for outer and immediate containers and to the instructions for use.
This document does not apply to:
a) IVD instruments or equipment;
b) IVD reagents for self-testing.

  • Standard
    24 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) instruments intended for self-testing.
This document is also applicable to apparatus and equipment intended to be used with IVD instruments for self-testing.
This document can also be applicable to accessories.
This document does not apply to:
a) instructions for instrument servicing or repair;
b) IVD reagents, including calibrators and control materials for use in control of the reagent;
c) IVD instruments for professional use.

  • Standard
    21 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) reagents, calibrators, and controls intended for self-testing.
This document can also be applicable to accessories.
This document is applicable to the labels for outer and immediate containers and to the instructions for use.
This document does not apply to:
a) IVD instruments or equipment;
b) IVD reagents for professional use.

  • Standard
    24 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document provides requirements for the development, validation and routine control of moist heat sterilization processes for medical devices. It also contains guidance which is intended to explain the requirements set forth in the normative sections. The guidance given is intended to promote good practice related to moist heat sterilization processes according to this document. The application within industrial and health care settings is considered.

  • Standard
    172 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document provides requirements for the development, validation and routine control of moist heat sterilization processes for medical devices. It also contains guidance which is intended to explain the requirements set forth in the normative sections. The guidance given is intended to promote good practice related to moist heat sterilization processes according to this document. The application within industrial and health care settings is considered.

  • Standard
    172 pages
    English language
    sale 10% off
    e-Library read for
    1 day

2021-11-09-JO-Following roundtable discussion on 2021-11-09 between EC desk officer, CCMC and CEN/TC 102, it was agreed that the assessment at ENQ will be performed by the Medical Device EC desk officer for a finale assessment at Enquiry. If positive, CEN/TC 102 may decide to skip the Formal Vote and proceed to publication.

  • Amendment
    33 pages
    English language
    sale 10% off
    e-Library read for
    1 day

2021-11-09-JO-Following roundtable discussion on 2021-11-09 between EC desk officer, CCMC and CEN/TC 102, it was agreed that the assessment at ENQ will be performed by the Medical Device EC desk officer for a finale assessment at Enquiry. If positive, CEN/TC 102 may decide to skip the Formal Vote and proceed to publication.

  • Amendment
    23 pages
    English language
    sale 10% off
    e-Library read for
    1 day

2021-11-09-JO-Following roundtable discussion on 2021-11-09 between EC desk officer, CCMC and CEN/TC 102, it was agreed that the assessment at ENQ will be performed by the Medical Device EC desk officer for a finale assessment at Enquiry. If positive, CEN/TC 102 may decide to skip the Formal Vote and proceed to publication.

  • Amendment
    33 pages
    English language
    sale 10% off
    e-Library read for
    1 day

2021-11-09-JO-Following roundtable discussion on 2021-11-09 between EC desk officer, CCMC and CEN/TC 102, it was agreed that the assessment at ENQ will be performed by the Medical Device EC desk officer for a finale assessment at Enquiry. If positive, CEN/TC 102 may decide to skip the Formal Vote and proceed to publication.

  • Amendment
    23 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document addresses good clinical practice for the design, conduct, recording and reporting of clinical investigations carried out in human subjects to assess the clinical performance or effectiveness and safety of medical devices.
For post-market clinical investigations, the principles set forth in this document are intended to be followed as far as relevant, considering the nature of the clinical investigation (see Annex I).
This document specifies general requirements intended to
—     protect the rights, safety and well-being of human subjects,
—     ensure the scientific conduct of the clinical investigation and the credibility of the clinical investigation results,
—     define the responsibilities of the sponsor and principal investigator, and
—     assist sponsors, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of medical devices.
NOTE 1  Users of this document need to consider whether other standards and/or national requirements also apply to the investigational device(s) under consideration or the clinical investigation. If differences in requirements exist, the most stringent apply.
NOTE 2  For Software as a Medical Device (SaMD) demonstration of the analytical validity (the SaMD's output is accurate for a given input), and where appropriate, the scientific validity (the SaMD's output is associated to the intended clinical condition/physiological state), and clinical performance (the SaMD's output yields a clinically meaningful association to the target use) of the SaMD, the requirements of this document apply as far as relevant (see Reference [4]). Justifications for exemptions from this document can consider the uniqueness of indirect contact between subjects and the SaMD.
This document does not apply to in vitro diagnostic medical devices. However, there can be situations, dependent on the device and national or regional requirements, where users of this document might consider whether specific sections and/or requirements of this document could be applicable.

  • Standard
    96 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document addresses good clinical practice for the design, conduct, recording and reporting of clinical investigations carried out in human subjects to assess the clinical performance or effectiveness and safety of medical devices.
For post-market clinical investigations, the principles set forth in this document are intended to be followed as far as relevant, considering the nature of the clinical investigation (see Annex I).
This document specifies general requirements intended to
—     protect the rights, safety and well-being of human subjects,
—     ensure the scientific conduct of the clinical investigation and the credibility of the clinical investigation results,
—     define the responsibilities of the sponsor and principal investigator, and
—     assist sponsors, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of medical devices.
NOTE 1  Users of this document need to consider whether other standards and/or national requirements also apply to the investigational device(s) under consideration or the clinical investigation. If differences in requirements exist, the most stringent apply.
NOTE 2  For Software as a Medical Device (SaMD) demonstration of the analytical validity (the SaMD's output is accurate for a given input), and where appropriate, the scientific validity (the SaMD's output is associated to the intended clinical condition/physiological state), and clinical performance (the SaMD's output yields a clinically meaningful association to the target use) of the SaMD, the requirements of this document apply as far as relevant (see Reference [4]). Justifications for exemptions from this document can consider the uniqueness of indirect contact between subjects and the SaMD.
This document does not apply to in vitro diagnostic medical devices. However, there can be situations, dependent on the device and national or regional requirements, where users of this document might consider whether specific sections and/or requirements of this document could be applicable.

  • Standard
    96 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.

  • Standard
    52 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    52 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.
It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.
It does not cover all requirements for packaging medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.

  • Standard
    21 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.

  • Standard
    52 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    52 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.
It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.
It does not cover all requirements for packaging medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.

  • Standard
    21 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 10993-4:2017 specifies general requirements for evaluating the interactions of medical devices with blood.
It describes
a)    a classification of medical devices that are intended for use in contact with blood, based on the intended use and duration of contact as defined in ISO 10993‑1,
b)    the fundamental principles governing the evaluation of the interaction of devices with blood,
c)    the rationale for structured selection of tests according to specific categories, together with the principles and scientific basis of these tests.
Detailed requirements for testing cannot be specified because of limitations in the knowledge and precision of tests for evaluating interactions of devices with blood. This document describes biological evaluation in general terms and may not necessarily provide sufficient guidance for test methods for a specific device.
The changes in this document do not indicate that testing conducted according to prior versions of this document is invalid. For marketed devices with a history of safe clinical use, additional testing according to this revision is not recommended.

  • Standard
    82 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the minimum functional and performance requirements for respiratory infection prevention devices (RIPDs).
RIPDs are intended to reduce the emission of infective agents from the user’s airways into the environment, and also reduce exposure to the user from inhalation of infective agents.
RIPDs are intended for use by everybody.

  • Draft
    22 pages
    English language
    sale 10% off
    e-Library read for
    1 day