ISO 20956:2023
(Main)Radiological protection — Low dose rate calibration of instruments for environmental and area monitoring
Radiological protection — Low dose rate calibration of instruments for environmental and area monitoring
This document specifies the calibration methods under laboratory conditions for dosemeters used for environmental and area monitoring of X and gamma-rays with respect to the operational quantities of the International Commission on Radiation Units and Measurements (ICRU)[1]. This document extends the dose rate range of ISO 4037-1 below 1,0 µSv·h−1. The specific uncertainty components are described for these calibration methods. This document also specifies the method for routine checking of active area dosemeters. Routine checking is not a calibration, nor does it replace a calibration, but it is a simple and effective method to routinely verify that the performance of the equipment is continuously maintained after calibration and that the calibration is still valid. This document does not deal with the special requirements for the calibration of spectrometer-based environmental dosemeters and passive dosemeters.
Radioprotection — Étalonnage d’instruments à faible débit de dose pour la surveillance de zone et de l’environnement
Le présent document spécifie les méthodes d’étalonnage dans des conditions de laboratoire pour les dosimètres utilisés pour la surveillance de zone et de l’environnement de rayons X et gamma en relation avec les grandeurs opérationnelles de la Commission internationale des unités et des mesures de radiation (ICRU)[1]. Le présent document étend la plage de débits de dose de l’ISO 4037-1 en dessous de 1,0 µSv·h−1. Les composantes d’incertitude spécifiques sont décrites pour ces méthodes d’étalonnage. Le présent document spécifie également la méthode de contrôle de routine des dosimètres de zone actifs. Le contrôle de routine n’est pas un étalonnage et ne remplace pas l’étalonnage; c’est une méthode simple et efficace qui permet de vérifier régulièrement que les performances de l’équipement sont maintenues sans interruption après l’étalonnage de sorte que l’étalonnage soit toujours valable. Le présent document ne traite pas des exigences spéciales relatives à l’étalonnage des dosimètres d’environnement utilisant un spectromètre et des dosimètres passifs.
General Information
Standards Content (Sample)
INTERNATIONAL ISO
STANDARD 20956
First edition
2023-09
Radiological protection — Low dose
rate calibration of instruments for
environmental and area monitoring
Radioprotection — Étalonnage d’instruments à faible débit de dose
pour la surveillance de zone et de l’environnement
Reference number
© ISO 2023
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Symbols . 3
5 Calibration methods under laboratory conditions . 3
5.1 Characterization of the radiation field using a reference source . 3
5.1.1 General . 3
5.1.2 Characterization procedure of the reference radiation field. 3
5.1.3 Characterization procedure of the radiation field at a distance r . 4
5.1.4 Uncertainty for the calibration of the radiation field using the reference
source . 4
5.2 Ground level facilities with normal background dose levels . 4
5.2.1 General . 4
5.2.2 Dose equivalent rate evaluation using the inverse square of distance . 4
5.2.3 Detector calibration procedure . 5
5.2.4 Uncertainty contributions to the detector calibration uncertainty . 5
5.3 Ground level facilities with added shielding at lower than normal background
levels . 5
5.3.1 General . 5
5.3.2 Description of the ground level facility with added shielding . 5
5.3.3 Detector calibration procedure . 5
5.3.4 Uncertainty contributions to the detector calibration uncertainty . 6
5.4 Underground facilities with ultra-low background dose levels . 6
5.4.1 General . 6
5.4.2 Description of the facility . 6
5.4.3 Uncertainty contributions to the detector calibration uncertainty . 6
6 Routine checking .7
6.1 General . 7
6.2 Description of the method . 7
6.2.1 Introduction . 7
6.2.2 Irradiation setup for regular checking . 7
6.2.3 Criteria for routine checks. 7
7 On-site calibration . 8
7.1 General . 8
7.2 The method using a portable calibrated radioactive source. 8
7.2.1 General . 8
7.2.2 Portable calibrated radioactive source . 8
7.2.3 Procedure of calibration . 8
7.2.4 Uncertainty contributions to the detector calibration uncertainty . 8
7.3 The method using a reference standard instrument . 8
7.3.1 General . 8
7.3.2 The calibration instruments . 9
7.3.3 Uncertainty contributions . 9
Annex A (informative) Example of a ground level facility with added shielding.10
Annex B (informative) Reference sites for characterization of environmental dosemeters
with respect to background ionizing radiation in the environment .12
Bibliography .14
iii
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO document should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use
of (a) patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed
patent rights in respect thereof. As of the date of publication of this document, ISO had not received
notice of (a) patent(s) which may be required to implement this document. However, implementers are
cautioned that this may not represent the latest information, which may be obtained from the patent
database available at www.iso.org/patents. ISO shall not be held responsible for identifying any or all
such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to
the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see
www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 85, Nuclear energy, nuclear technologies
and radiological protection, Subcommittee SC 2, Radiological protection.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
iv
Introduction
The ISO 4037 documentary standard is an International Standard for the calibration of dosemeters in
X and gamma-ray reference fields. The air kerma rate range for which the standard applies is above
−1 −1
1 µGy·h being nearly equivalent to 1 µSv·h in terms of the operational quantities. The standard
however does not address dosemeters that are currently available for environmental monitoring that
−1
can measure below 1 µSv·h . The reliability of environmental monitoring equipment is important for
the evaluation of potential dose of the general public during any radiation incident such as an accident
that could occur at a nuclear power plant or at any radiation facility.
This document extends the scope of ISO 4037 for calibrating area and environmental type monitors to
−1
dose rates below 1 µSv·h to improve the reliability of environmental monitoring.
Three detector calibration methods are described in Clause 5 of this document for 3 different situations:
ground level facilities with normal background levels (see 5.2), ground level facilities with lower than
normal background levels (see 5.3) and underground facilities with lower than normal background
levels (see 5.4). Clause 6 discusses routine checks while Clause 7 is dedicated to the calibration of
environmental and area dosemeters that are fixed in place. These methods are based on free-in-air
and simultaneous irradiation procedures, derived from those described in ISO 29661. This document
−1
extends the dose rate range of ISO 4037-1 below 1,0 µSv·h . The specific uncertainty components are
described for these calibration methods.
v
INTERNATIONAL STANDARD ISO 20956:2023(E)
Radiological protection — Low dose rate calibration of
instruments for environmental and area monitoring
1 Scope
This document specifies the calibration methods under laboratory conditions for dosemeters used for
environmental and area monitoring of X and gamma-rays with respect to the operational quantities of
[1]
the International Commission on Radiation Units and Measurements (ICRU) .
−1
This document extends the dose rate range of ISO 4037-1 below 1,0 µSv·h . The specific uncertainty
components are described for these calibration methods.
This document also specifies the method for routine checking of active area dosemeters. Routine
checking is not a calibration, nor does it replace a calibration, but it is a simple and effective method
to routinely verify that the performance of the equipment is continuously maintained after calibration
and that the calibration is still valid.
This document does not deal with the special requirements for the calibration of spectrometer-based
environmental dosemeters and passive dosemeters.
2 Normative references
The following documents are refe
...
NORME ISO
INTERNATIONALE 20956
Première édition
2023-09
Radioprotection — Étalonnage
d’instruments à faible débit de dose
pour la surveillance de zone et de
l’environnement
Radiological protection — Low dose rate calibration of instruments
for environmental and area monitoring
Numéro de référence
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2023
Tous droits réservés. Sauf prescription différente ou nécessité dans le contexte de sa mise en œuvre, aucune partie de cette
publication ne peut être reproduite ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique,
y compris la photocopie, ou la diffusion sur l’internet ou sur un intranet, sans autorisation écrite préalable. Une autorisation peut
être demandée à l’ISO à l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Genève
Tél.: +41 22 749 01 11
E-mail: copyright@iso.org
Web: www.iso.org
Publié en Suisse
ii
Sommaire Page
Avant-propos .v
Introduction . vi
1 Domaine d’application . 1
2 Références normatives .1
3 Termes et définitions . 1
4 Symboles . 3
5 Méthodes d’étalonnage dans des conditions de laboratoire . 3
5.1 Caractérisation du champ de rayonnement à l’aide d’une source de référence . 3
5.1.1 Généralités . 3
5.1.2 Mode opératoire de caractérisation du champ de rayonnement de référence . 3
5.1.3 Mode opératoire de caractérisation du champ de rayonnement à une
distance r . 4
5.1.4 Incertitude de l’étalonnage du champ de rayonnement à l’aide de la source
de référence . 4
5.2 Installations au niveau du sol avec des niveaux de bruit de fond habituels . 4
5.2.1 Généralités . 4
5.2.2 Évaluation du débit d’équivalent de dose en utilisant l’inverse du carré de
la distance . 5
5.2.3 Mode opératoire d’étalonnage du détecteur . 5
5.2.4 Contributions à l’incertitude d’étalonnage du détecteur . 5
5.3 Installations au niveau du sol munies d’un blindage supplémentaire à des niveaux
de bruit de fond inférieurs à la normale . 5
5.3.1 Généralités . 5
5.3.2 Description de l’installation au niveau du sol avec blindage supplémentaire . 5
5.3.3 Mode opératoire d’étalonnage du détecteur . 6
5.3.4 Contributions à l’incertitude d’étalonnage du détecteur . 6
5.4 Installations souterraines avec des niveaux de bruit de fond ultra-faibles . 6
5.4.1 Généralités . 6
5.4.2 Description de l’installation . 6
5.4.3 Contributions à l’incertitude d’étalonnage du détecteur . 7
6 Contrôle de routine . 7
6.1 Généralités . 7
6.2 Description de la méthode . 7
6.2.1 Introduction . 7
6.2.2 Configuration de l’irradiation pour contrôle régulier . 7
6.2.3 Critères pour les contrôles de routine . 8
7 Étalonnage sur site .8
7.1 Généralités . 8
7.2 Méthode utilisant une source radioactive portable étalonnée . 8
7.2.1 Généralités . 8
7.2.2 Source radioactive portable étalonnée . 8
7.2.3 Mode opératoire d’étalonnage . . 9
7.2.4 Contributions à l’incertitude d’étalonnage du détecteur . 9
7.3 Méthode utilisant un instrument étalon de référence . 9
7.3.1 Généralités . 9
7.3.2 Les instruments d’étalonnage . 9
7.3.3 Contribution à l’incertitude . 9
Annexe A (informative) Exemple d’installation au niveau du sol équipée d’un blindage
supplémentaire .10
Annexe B (informative) Sites de référence pour la caractérisation de dosimètres
d’environnement par rapport au bruit de fond du rayonnement ionisant ambiant .12
iii
Bibliographie .14
iv
Avant-propos
L’ISO (Organisation internationale de normalisation) est une fédération mondiale d’organismes
nationaux de normalisation (comités membres de l’ISO). L’élaboration des Normes internationales est
en général confiée aux comités techniques de l’ISO. Chaque comité membre intéressé par une étude
a le droit de faire partie du comité technique créé à cet effet. Les organisations internationales,
gouvernementales et non gouvernementales, en liaison avec l’ISO participent également aux travaux.
L’ISO collabore étroitement avec la Commission électrotechnique internationale (IEC) en ce qui
concerne la normalisation électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier, de prendre note des différents
critères d’approbation requis pour les différents types de documents ISO. Le présent document
a été rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2
(voir www.iso.org/directives).
L’ISO attire l’attention sur le fait que la mise en application du présent document peut entraîner
l’utilisation d’un ou de plusieurs brevets. L’ISO ne prend pas position quant à la preuve, à la validité et
à l’applicabilité de tout droit de propriété revendiqué à cet égard. À la date de publication du présent
document, l’ISO n’avait pas reçu notification qu’un ou plusieurs brevets pouvaient être nécessaires à sa
mise en application. Toutefois, il y a lieu d’avertir les responsables de la mise en application du présent
document que des informations plus récentes sont susceptibles de figurer dans la base de données de
brevets, disponible à l’adresse www.iso.org/brevets. L’ISO ne saurait être tenue pour responsable de ne
pas avoir identifié de tels droits de brevets.
Les appellations commerciales éventuellement mentionnées dans le présent document sont données
pour information, par souci de commodité, à l’intention des utilisateurs et ne sauraient constituer un
engagement.
Pour une explication de la nature volontaire des normes, la signification des termes et expressions
spécifiques de l’ISO liés à l’évaluation de la conformité, ou pour toute information au sujet de l’adhésion
de l’ISO aux principes de l’Organisation mondiale du commerce (OMC) concernant les obstacles
techniques au commerce (OTC), voir www.iso.org/avant-propos.
Le présent document a été préparé par le comité technique ISO/TC 85, Énergie nucléaire,
technologies nucléaires, et radioprotection, sous-comité SC 2, Radioprotection.
Il convient que l’utilisateur adresse tout retour d’information ou toute question concernant le présent
document à l’organisme national de normalisation de son pays. Une liste exhaustive desdits organismes
se trouve à l’adresse www.iso.org/fr/members.html.
v
Introduction
La norme documentaire ISO 4037 est une Norme internationale pour l’étalonnage des dosimètres dans
les champs de référence de rayonnements X et gamma. La plage de débits de kerma dans l’air à laquelle
−1 −1
la norme s’applique est supérieure à 1 µGy·h , ce qui est presque équivalent à 1 µSv·h en termes de
grandeurs opérationnelles. La norme ne couvre cependant pas les dosimètres aujourd’hui disponibles
−1
pour la surveillance de l’environnement qui peuvent mesurer à moins de 1 µSv·h . La fiabilité des
équipements de surveillance de l’environnement est importante pour l’évaluation de la dose d’exposition
potentielle du public en cas d'accident impliquant des rayonnements, par exemple, à un accident qui se
produirait dans une centrale nucléaire ou toute installation émettant des rayonnements.
Le présent document étend le domaine d’application de l’ISO 4037 pour l’étalonnage de moniteurs de
−1
zone et de l’environnement à des débits de dose inférieurs à 1 µSv·h afin d’améliorer la fiabilité de la
surveillance de l’environnement.
Trois méthodes d’étalonnage de détecteur sont décrites à l’Article 5 du présent document pour
3 situations différentes: les installations au niveau du sol avec des niveaux de bruit de fond normaux
(voir 5.2), les installations au niveau du sol avec des niveaux de bruit de fond inférieurs à la normale
(voir 5.3) et les installations souterraines avec des niveaux de bruit de fond inférieurs à la normale
(voir 5.4). L’Article 6 traite des contrôles de routine tandis que l’Article 7 est consacré à l’étalonnage des
dosimètres de zone et d’environnement installés à poste fixe. Ces méthodes sont fondées sur des modes
opératoires d’irradiations dans l’air en champ non perturbé et simultanées, dérivés de ceux décrits
dans l’ISO 29661. Le présent document étend la plage de débits de dose de l’ISO 4037-1 en dessous de
−1
1,0 µSv·h . Les composantes d’incertitude spécifiques sont décrites pour ces méthodes d’étalonnage.
vi